主要特征
可測量高達 92% 的高濃度磷酸
無需冷卻機制和冷卻時間。
無需冷卻循環(huán)管路中的高溫磷酸(140 至 170℃),可直接測量。
樣品使用PFA接觸,以降低污染風(fēng)險。
采用每半年一次的背景校正周期,有助于顯著減少機臺停機時間。
每3秒更新一次測量數(shù)據(jù),有助于更快速使用濃度反饋進行控制
穩(wěn)定性
? 測量結(jié)果
樣品(實際藥液)測量 1 小時,展示出良好的穩(wěn)定性。
? 測量條件
測量周期:每3秒 測量時間:1 小時 移動平均數(shù):16 次
樣品濃度(初始樣品濃度)H3PO4:92.2% H2O: 7.8% 樣品溫度:141°C
Product name
|
CHEMICAL SOLUTION MONITOR
|
Model
|
CS-620F
|
Measurement principle
|
Absorption spectroscopic method
|
Calculation principle
|
Temperature compensation type multivariate analysis
|
Sample
|
H3PO4/H2O
|
Measurable range
Repeatability
|
|
Component
|
Measurable range
(mass%)
|
Repeatability
(mass%)
|
Analog output
range(mass%)
|
1st solution
|
H3PO4
|
85.00-92.00
|
+/-0.10
|
85.00-100.00
|
H2O
|
8.0-15.0
|
+/-0.3
|
0.0-15.0
|
? Repeatability is defined by maximum error from average (1 hour).
? In the case of low temperature measurement (140°C) and high temperature measurement (170°C) with the same chemical,
there is possibility of discrepancy (Maximum +/- 0.30%).
|
Conditions
of Measurement
|
1) Measurement interval: Approx. 3 sec. (minimum)
2) Moving average: 16 times
|
Connection fitting size
|
1 inch or 3/4 inch
|
Sample condition
|
Sample temperature: 140 to 170°C
Atmospheric temperature, Chemical temperature fluctuation: within +/- 1°C (period of time : 1 hour)
Input pressure: 0.2 MPa or less
Pressure fluctuation: 0.02 MPa or less
Flow rate: 1 to 30 L/min
|
Air(for operation and purge)
|
Connection port: 4 mm O.D. quick joint
Pressure: 0.2 MPa ± 0.02 MPa
|
Power source
|
100 to 230 V AC (Single-Phase), 50/60 Hz
|
Power consumption
|
Approx.85 VA (Transient electric current at the time of the start is excluded)
|
Communication
|
Parallel I/O, RS-232C, Analog output
|
Dimension
|
(Monitor) 200 (W) × 308 (D) × 78 (H) mm
(Lamp unit) 200 (W) × 262 (D) × 100 (H) mm
(Sample cell) 240 (W) × 209 (D) × 205 (H) mm
|
Weight
|
(Monitor) Approx.3.6 kg
(Lamp unit) Approx.2.8 kg
(Sample cell) Approx.3.0 kg
|
Ambient temperature
|
(Monitor, Lamp unit) 20 to 30°C
(Optical fiber, Sample cell) 20 to 100°C
* Sudden temperature change should be avoided, within +/- 1°C/1hour
|
Ambient humidity
|
(Monitor, Lamp unit) 40 to 70% (Should be no dew condensation)
|
Slanting angle of installation
|
(Monitor, Lamp unit) Within +/-1 degree
(Sample cell) Please install so that air bubbles do not stay up.
|
Optical fiber
|
Length: 5 m, Bend radius: R150 mm
|